# tinySA hardware validation checklist This checklist qualifies one physical instrument and firmware build. A passing mock or CI run is not hardware evidence. Run the validator once with a tinySA Basic and once with a tinySA Ultra; keep their reports separate. ## Before each run - [ ] Identify the rear-label model and record the firmware shown by the device. - [ ] Disconnect every RF source. This test needs only the analyzer and a data-capable USB cable. - [ ] Close serial terminals and other tinySA software. Confirm the current user can open the USB serial port without running the validator as root. - [ ] If the front-panel configuration matters, save a preset. The validator changes repeat, raw zero offset, input mode, sweep limits, and pause state. From the repository root, run exactly one command: ```console scripts/test-hardware.sh basic ``` For an Ultra, use: ```console scripts/test-hardware.sh ultra ``` The script asks when to unplug and plug the instrument so it can detect the new serial port. If discovery is ambiguous, choose from the displayed ports. To skip discovery, pass `--port /dev/ttyACM0` or `--port COM7`. It builds the validator, runs two 101-point scans, validates the returned values, writes CSV files and an identity report under ignored `hardware-validation/`, and asks whether the front-panel sweep resumed. ## Basic acceptance - [ ] The automatically selected port belongs to the Basic. - [ ] Both `legacy` (240–300 MHz low input) and `direct` (350–400 MHz high input) scans report `PASS` and create CSV files with 101 data rows plus a units header. - [ ] The printed `version`/`info` text matches the connected Basic and its recorded firmware. - [ ] The live front-panel sweep resumes after the scans; answer `y` only after checking it. - [ ] The final `basic-report.txt` records both CSV paths and `operator_confirmed=y` (or `yes`). ## Ultra acceptance and limit - [ ] The automatically selected port belongs to the Ultra. - [ ] Both `legacy` and `direct` scans report `PASS` at 240–300 MHz and create CSV files with 101 data rows plus a units header. - [ ] The printed `version`/`info` text identifies the exact connected Ultra variant and firmware. - [ ] The live front-panel sweep resumes after the scans; answer `y` only after checking it. - [ ] The final `ultra-report.txt` records both CSV paths and `operator_confirmed=y` (or `yes`). A passing Ultra run validates only the shared low-input command path below 350 MHz, including `zero 128`, binary framing, CSV export, and best-effort resume. It does not validate Ultra mode, harmonic mode, LNA behavior, ranges at or above 350 MHz, level accuracy, or frequency accuracy. Those require model-aware acquisition and a controlled reference source; do not infer them from this communications test. ## Failure record - [ ] Keep the generated report and any completed CSV; do not edit them. - [ ] Record the command, host OS, selected port, exact model, firmware, and the complete `FAIL:` line. - [ ] If the sweep did not resume, disconnect any RF setup and use `resume`, reload the saved preset, or power-cycle the instrument. - [ ] Do not connect a transmitter merely to reproduce a communications failure.