tinySA hardware validation checklist

This checklist qualifies one physical instrument and firmware build. A passing mock or CI run is not hardware evidence. Run the validator once with a tinySA Basic and once with each supported tinySA Ultra profile; keep reports separate.

Before each run

  • Identify the rear-label model and record the firmware shown by the device.
  • Disconnect every RF source. This test needs only the analyzer and a data-capable USB cable.
  • Close serial terminals and other tinySA software. Confirm the current user can open the USB serial port without running the validator as root.
  • If the front-panel configuration matters, save a preset. The validator changes repeat, raw zero offset, input mode, sweep limits, and pause state. The ZS406 profile also enables Ultra mode, which firmware may retain.

From the repository root, run exactly one command:

scripts/test-hardware.sh basic

For an Ultra, use:

scripts/test-hardware.sh ultra

For the ZS406 normal and non-harmonic Ultra ranges, use:

scripts/test-hardware.sh zs406

The script asks when to unplug and plug the instrument so it can detect the new serial port. If discovery is ambiguous, choose from the displayed ports. To skip discovery, pass --port /dev/ttyACM0 or --port COM7. It builds the validator, runs two 101-point scans, validates the returned values, writes CSV files and an identity report under ignored hardware-validation/, and asks whether the front-panel sweep resumed.

Basic acceptance

  • The automatically selected port belongs to the Basic.
  • Both legacy (240–300 MHz low input) and direct (350–400 MHz high input) scans report PASS and create CSV files with 101 data rows plus a units header.
  • The printed version/info text matches the connected Basic and its recorded firmware.
  • The live front-panel sweep resumes after the scans; answer y only after checking it.
  • The final basic-report.txt records both CSV paths and operator_confirmed=y (or yes).
  • Each report entry records raw_level_offset_db=128.

Recorded Basic evidence

The Basic acquisition path was validated on physical hardware on 2026-10-05 with firmware tinySA_v1.4-175-g1419a93. Both the 240–300 MHz low-input scan and the 350–400 MHz high-input scan returned 101 valid samples, exercised the fixed-record binary decoder, produced CSV output, and resumed the live sweep. This is communications and workflow evidence, not frequency or level-accuracy calibration.

Ultra acceptance and limit

  • The automatically selected port belongs to the Ultra.
  • Both legacy and direct scans report PASS at 240–300 MHz and create CSV files with 101 data rows plus a units header.
  • The printed version/info text identifies the exact connected Ultra variant and firmware.
  • The live front-panel sweep resumes after the scans; answer y only after checking it.
  • The final ultra-report.txt records both CSV paths and operator_confirmed=y (or yes).
  • Each report entry records raw_level_offset_db=174.

A passing Ultra run validates only the shared low-input command path below 350 MHz, including the native zero 174 encoding, binary framing, CSV export, and best-effort resume. It does not validate Ultra mode, harmonic mode, LNA behavior, ranges at or above 350 MHz, level accuracy, or frequency accuracy. Those require model-aware acquisition and a controlled reference source; do not infer them from this communications test.

ZS406 range acceptance

  • The rear label says ZS406 and identity reports ZS406 or hardware V0.4.6.
  • normal (0.1–900 MHz) and ultra (900 MHz–5.4 GHz) each report PASS and create 101-row CSV files with units headers.
  • The report records enable_ultra_mode=false for normal, true for ultra, and raw_level_offset_db=174 for both.
  • The live front-panel sweep resumes and the final zs406-report.txt records affirmative operator confirmation.

This validates identity, accepted sweep bounds, command ordering, binary framing and finite decoded values across both ranges. With no controlled RF source it does not establish frequency accuracy, level accuracy, sensitivity, spur rejection or the authenticity of displayed signals.

Recorded ZS406 evidence

The ZS406 acquisition profile was validated on physical hardware on 2026-10-07 with hardware V0.4.6 and firmware tinySA4_v1.4-197-gaa78ccc. The 100 kHz–900 MHz normal scan and 900 MHz–5.4 GHz Ultra scan each returned 101 finite samples, confirmed the requested sweep boundaries, produced CSV output, and resumed the front-panel sweep. The RF input was unconnected. This is communications, range-acceptance and workflow evidence, not RF frequency or level calibration.

Failure record

  • Keep the generated report and any completed CSV; do not edit them.
  • Record the command, host OS, selected port, exact model, firmware, and the complete FAIL: line.
  • If the sweep did not resume, disconnect any RF setup and use resume, reload the saved preset, or power-cycle the instrument.
  • Do not connect a transmitter merely to reproduce a communications failure.