Basic and Ultra acquisition support

Implemented 2026-10-05; synthetic verification is separate from hardware qualification.

The acquisition path reads version, info, and help before changing scan settings. Recognized Basic (tinySA_v…) and Ultra-family (tinySA4_v…) identities select different plans. Unknown or conflicting identities fail. Explicit scan.device_model = "basic" or "ultra" requires a matching detected family; the default is "auto". Family detection does not identify a hardware revision or certify that an arbitrary firmware build implements every command correctly.

Family Implemented acquisition range Input command Automatic raw bias
Basic Existing low/high policy through 960 MHz mode low input or mode high input 128 dB
Ultra/Ultra+ Common normal range 0.1–800 MHz mode low input 174 dB

The official comparison lists 800 MHz normal range for ZS405 and 900 MHz for Ultra+ revisions. This initial implementation deliberately uses their shared 800 MHz ceiling. Extended/Ultra/harmonic ranges, revision detection, LNA control and mode-state restoration are not implemented. Existing instrument processing settings and Ultra-enable state are retained; no unlock or persistent-save command is sent. The official Ultra console snapshot documents the family version prefix and zero readback. The reviewed mode handler accepts mode low input for Ultra. Sources checked 2026-10-05.

Required commands must appear in help. Optional diagnostics run only when advertised. After setting the selected bias with zero N, acquisition queries zero and requires a matching NdBm readback. A missing or mismatched readback fails before scanning. The decoder uses this confirmed value. Both families share the same binary record layout; see protocol.

Configuration migration

New defaults select device_model = "auto", auto_zero_offset = true, and 240–800 MHz, usable by both families. Saved frequencies remain unchanged; an old 960 MHz stop is rejected on Ultra with an actionable range error. To preserve a custom raw bias, set auto_zero_offset = false alongside the existing zero_offset value. Previously saved offsets remain parseable but are ignored while automatic selection is enabled. No configuration is rewritten until the normal save workflow runs.

The terminal uses integer MHz inputs and a shared 960 MHz entry ceiling. Model-specific validation runs after identification in the acquisition worker; therefore out-of-profile input is rejected before mutation, but after the UI saves the selection. Use the library for sub-MHz settings. Dynamic UI bounds and detection before the selection screen remain future work.

Verification and qualification

Regression tests cover model detection, conflicts, range/path selection, automatic native offsets, command inventory, readback failures, partial I/O and resume after scan failure. Run the local gates and cargo coverage. The hardware validator now requires its selected family to match detection before changing settings and exercises Ultra below 800 MHz.

This is initial software support, not a claim of hardware-qualified first-tier support. Existing Basic evidence predates the new help/zero readback flow; rerun hardware validation on both instruments. Ultra RF level/frequency accuracy, Windows/macOS, Jenkins and publication have not been verified by these synthetic tests.