tinySA hardware validation checklist
This checklist qualifies one physical instrument and firmware build. A passing mock or CI run is not hardware evidence. Run the validator once with a tinySA Basic and once with a tinySA Ultra; keep their reports separate.
Before each run
- Identify the rear-label model and record the firmware shown by the device.
- Disconnect every RF source. This test needs only the analyzer and a data-capable USB cable.
- Close serial terminals and other tinySA software. Confirm the current user can open the USB serial port without running the validator as root.
- If the front-panel configuration matters, save a preset. The validator changes repeat, raw zero offset, input mode, sweep limits, and pause state.
From the repository root, run exactly one command:
scripts/test-hardware.sh basic
For an Ultra, use:
scripts/test-hardware.sh ultra
The script asks when to unplug and plug the instrument so it can detect the new
serial port. If discovery is ambiguous, choose from the displayed ports. To skip
discovery, pass --port /dev/ttyACM0 or --port COM7. It builds the validator,
runs two 101-point scans, validates the returned values, writes CSV files and an
identity report under ignored hardware-validation/, and asks whether the
front-panel sweep resumed.
Basic acceptance
- The automatically selected port belongs to the Basic.
- Both
legacy(240–300 MHz low input) anddirect(350–400 MHz high input) scans reportPASSand create CSV files with 101 data rows plus a units header. - The printed
version/infotext matches the connected Basic and its recorded firmware. - The live front-panel sweep resumes after the scans; answer
yonly after checking it. - The final
basic-report.txtrecords both CSV paths andoperator_confirmed=y(oryes). - Each report entry records
raw_level_offset_db=128.
Ultra acceptance and limit
- The automatically selected port belongs to the Ultra.
- Both
legacyanddirectscans reportPASSat 240–300 MHz and create CSV files with 101 data rows plus a units header. - The printed
version/infotext identifies the exact connected Ultra variant and firmware. - The live front-panel sweep resumes after the scans; answer
yonly after checking it. - The final
ultra-report.txtrecords both CSV paths andoperator_confirmed=y(oryes). - Each report entry records
raw_level_offset_db=174.
A passing Ultra run validates only the shared low-input command path below
350 MHz, including the native zero 174 encoding, binary framing, CSV export,
and best-effort
resume. It does not validate Ultra mode, harmonic mode, LNA behavior, ranges at
or above 350 MHz, level accuracy, or frequency accuracy. Those require
model-aware acquisition and a controlled reference source; do not infer them
from this communications test.
Failure record
- Keep the generated report and any completed CSV; do not edit them.
- Record the command, host OS, selected port, exact model, firmware, and the
complete
FAIL:line. - If the sweep did not resume, disconnect any RF setup and use
resume, reload the saved preset, or power-cycle the instrument. - Do not connect a transmitter merely to reproduce a communications failure.